| Predictive
  Accuracy of Microvolt T-Wave Alternans Testing in Primary Prevention Patients
  With and Without ICDs | |
| Presentation Start/End Time: | Thursday, May 15, 2008, 1:30 PM - 1:45
  PM | 
| Author Block: | Stefan H.
  Hohnloser, MD, Takanori Ikeda, MD, PhD and Richard J. Cohen, MD, PhD. J. W.
  Goethe University, Frankfurt, Germany, Kyorin University School of Medicine,
  Tokyo, Japan, Massachusetts Institute of Technology, Cambridge, MA | 
| Introduction:
  Microvolt T-wave alternans (MTWA) testing has proven in many studies to be a
  highly accurate predictor of ventricular tachyarrhythmic events (VTEs) in
  patients with risk factors for sudden cardiac death but without a prior
  history of sustained VTEs (primary prevention patients). In some recent
  studies MTWA has not performed as well. In the present analysis we examine
  the hypothesis that MTWA is an accurate predictor of VTEs in patients without
  implanted ICDs, but not of appropriate ICD therapy (AICD-Rx) in patients with
  prophylactically implanted ICDs. Methods: We identified prospective clinical
  trials evaluating MTWA measured using the spectral analytic method in primary
  prevention populations and analyzed studies in which (1) few patients had
  implanted ICDs and <= 15% of reported endpoint VTEs were AICD-Rx or (2) a
  majority of reported endpoint VTEs were AICD-Rx. Results: In the low AICD-Rx
  group comprising 3,682 patients the hazard ratio associated with a
  non-negative vs negative MTWA test was 13.6 [8.5-30.4] (brackets denote 95%
  confidence interval, p < 0.0001), and the annual event rate (AER) among
  MTWA negative patients was 0.3% [0.1%-0.5%]. In contrast, in the high AICD-Rx
  group comprising 2,234 patients the hazard ratio was only 1.6 [1.2-2.1] (p
  < 0.001), and the AER among MTWA negative patients was elevated to 5.4%
  [4.1%-6.7%]. In support of these findings we analyzed published data from the
  MADIT II and SCD-HeFT trials and determined that in those trials only 32% of
  patients who received AICD-Rx averted SCD. Conclusions: MTWA testing provides
  an accurate means of predicting SCD in primary prevention patients without
  implanted ICDs; in particular, the event rate is very low among patients with
  a negative MTWA test. In prospective trials of ICD therapy the number of
  AICD-Rx’s greatly exceed the number of SCDs that would have occurred in the
  absence of ICD implantation. In trials involving patients with implanted ICDs
  these excess AICD-Rx’s appear to distribute randomly between MTWA negative
  and non-negative patients obscuring the predictive accuracy of MTWA for SCD. AICD-Rx
  is an unreliable surrogate for SCD in clinical trials. | |